Aligning the AFM Laser onto the Cantilever

Most commercial Atomic Force Microscopes (AFMs) utilize a light-lever force sensor which reflects a focused laser from the back side of a cantilever. The cantilevers can be 20-30 microns wide by 50-300 microns long. The cantilever, cantilever chip, and the probe work together as a coordinated unit, and are micro-machined.

Some AFMs offer 'pre-aligned' cantilever/cantilever chip/probe units, theoretically eliminating the task of properly landing and reflecting the laser on the back side of the cantilever. The pre-aligned concept ends up being a challenge to achieve due to the impacts of contamination, the subtle variations in cantilever dimensions, and human error. Thus, even with a 'pre-aligned' design, it becomes necessary to include a mechanical adjustment mechanism on the AFM so that the user can physically move the focused laser beam onto the cantilever. An optical microscope makes a big difference when the AFM user, whether working with a pre-aligned cantilever or a user-aligned cantilever, manually positions the focused laser properly onto the cantilever's back side.  The two images at the top of the page help illustrate this point.

Locating Sample Surface Features for AFM Scanning

Atomic Force Microscopes typically have scan ranges that are less than 100 microns. This means it's not possible to view large surface areas - or to "get the big picture". That makes sense when you consider you're trying to visualize very small objects, but the problem becomes trying to 'find' the important areas of the sample. For example, the area of interest may be on isolated areas of the sample. A high-quality optical microscope enables the user to rapidly make determinations about the areas of a sample to scan with the AFM. Having a large optical zoom ratio is really helpful: at low magnification the user can find the general area for scanning, and at high magnification the user can find specific areas for AFM scanning.

test structure sample viewed by optical microscope for AFM scanning

In the photos above, the left image shows a low magnification view from the optical microscope of a test structure. On the very right hand side of the image, the AFM probe/cantilever is visible.  A large view like that on the left facilitates finding a particular feature for AFM scanning. The right image shows a higher magnification optical microscope view of the same test structure.  This image demonstrates how, once the user has found a particular feature for scanning, the user can then move the AFM probe to a position just above the region to be scanned. (Note: the optical microscope is focused on the sample surface; as the probe is above the surface, it is not in focus.)

Conducting Successful Probe Approach

In an Atomic Force Microscope (AFM) the probe tip must be moved to a point where it interacts with the various forces associated with a sample's surface. To make the tip's approach to the sample faster, it is helpful to get the probe as close to the surface as possible before utilizing an automated tip approach algorithm. This probe approach is facilitated with an optical microscope in two ways:
Focusing the optical microscope just above the sample surface, and then using the AFM Z approach mechanism to bring the probe closer until it is focused, i.e. very close to the surface.

proper laser alignment on a AFM probe

Above, the left view from an optical microscope shows that when the probe is away from the sample surface, the laser's spot is a distance from the cantilever. Additionally, when the probe is not on the surface, only the sample or the cantilever will be in focus. The right view shows that when the probe reaches the sample surface, both the cantilever and the sample are in focus, and the two laser spots overlap.

In AFMs with a laser beam that is not at a perpendicular angle to the sample's surface, the motion of the laser spot moving across the surface is used to facilitate probe approach. As the tip gets closer to the surface, the two laser spots will coalesce. 

Three Viewing Orientations for Optical and Atomic Force Microscope Combinations

The viewing orientations of the optical microscope in relation to the AFM probe, cantilever, and sample are:

  • top view
  • bottom view
  • 45 degree angle view

The top view is the most popular optical microscope orientation with the AFM. The optical microscope is placed directly above the AFM cantilever, yielding an optical view of the top of the AFM cantilever and the top of the surface of the sample to be imaged. Typically, a uniaxial light source is required for a top view microscope.

TT-AFM optics and overhead sample view

Above, the left image shows a TT-AFM stage with top view optical microscope with zoom tube, xy translator, focus control, and CMOS camera. The right image shows a top-view of a probe in close proximity to a feature of interest.

The bottom view is possible with an inverted light microscope (ILM). In this configuration, the user can view the sample and cantilever from the bottom. This type of view is very helpful for life sciences applications, and is popular for other types of microscopy such as flourescence microscopy.

inverted light microscope and AFM probe plus cell sample

Left, an inverted light microscope's view of the AFM probe. Right, an inverted light microscope's image of a cell and an AFM cantilever, with an area of interest highlighted for AFM scanning.

The 45 degree angle view can be helpful for probe approach, but it is not necessary if the optical microscope's top-view optic has a decent focus control.

Optical, inverted, and AFM view of e. coli sample

The above image shows three views of the same e. coli sample, from AFMWorkshop's LS-AFM: L to R: video optical microscope, inverted optical microscope, and atomic force microscope. The area in red indicates sample area selected for AFM scanning.

chart of minimum optical microscope specs for AFM use

×